ICAN: Ion Slicer

Ion Slicer

The Ion Slicer is used for the preparation of thin films using argon ion sputtering.

With this method samples for TEM-,STEM-,SEM- or SPM-investigations can be fabricated without the need for any chemicals.

The argon ions have a small incident angly onto the sample surface, minimizing radiation damage and amorphisation.

The sample can be tiltes by 30°, to achieve a homogenously polished area.

JEOL Ion Slicer

Specifications:

Acceleration voltage: 1-8 kV
Sputter rate: 5µm/min
Sample size: 0.8 mm (B )x 2.8 mm (L)x 0.1 mm (H)
Sample rotation: +/-30°
Gas für ion beam: Argon
Incident angle: 0°-6°
Pressure gauge: Penning tube
Main pump: turbomolecular pump(TMP)
Pre pump: rotary vane pump

Ican Logo Rgb 230x195

Cenide Service